After-sales guidance


Breakup

Replace the thermocouple.

The SV and PV values on the instrument display are garbled, flashing continuously, and the furnace status is not shown.

Thermocouple failure or loose screws

Turn on the LOCK switch; the PV displays code 1213.

Thermocouple is faulty.

As soon as I turn it on, the temperature starts fluctuating wildly.

Thermocouple is faulty.

The instrument panel displays error code “HIAL”.

Thermocouple loose; tighten it to resolve the issue.

BTF-1200C-III-S-S Third Temperature Zone Instrument Display: 1381

Thermocouple is faulty.

The temperature was set to 700°C, and after maintaining that temperature for a while, the instrument displayed 1000°C. We suspect there is a problem with the temperature control system.

Preliminary diagnosis indicates a broken thermocouple. Replace the thermocouple.

The vacuum can’t be drawn down.

After power is applied, the motor does not rotate; replace the vacuum pump.

The vacuum can’t be drawn down.

The shut-off valve was installed in the wrong direction. The issue has been resolved by following the correct installation procedure.

The vacuum can’t be drawn down.

Replace the pump oil.

The vacuum hot-pressing equipment is broken; the pressure won’t build up.

The check valve is stuck and damaged.

The tube furnace stops pumping down at 150 Pa and cannot achieve a lower vacuum.

After removing the dust filter, the vacuum pump was run until a pressure of 0.1 Pa was achieved, and then the dust filter was replaced.

Vacuum issues with tube furnaces

Replace the faulty manual slide valve (KF25, 1 unit).

The resistance window does not display a number.

The resistance gauge has failed; replace the resistance gauge.

The vacuum can’t be drawn down.

After installing the sealing ring cavity, the vacuum level is normal.

The vacuum can’t be drawn down.

The flange was not installed properly; remove it and reinstall it, after which you can perform a vacuum draw.

The vacuum can’t be drawn down.

The manual valve was not fully opened; after opening it, the vacuum pump operates normally.

BTF-1200C: After disassembling the flange to clean the inside of the quartz tube and reassembling it, the vacuum level can no longer reach 0.2 Pa; currently, the highest achievable vacuum is only 13 Pa.

Replace the sealing ring and the pump oil.

The vacuum can’t be pulled down.

The mechanical pump’s vacuum level, measured under single‑test conditions, is 11 Pa. To determine whether the mechanical pump is contaminated, it is recommended to send it for repair or replacement.

The molecular pump controller has no power.

The molecular pump controller fuse has blown; replace it with a fuse of the same specification. The molecular pump is now operating normally.