I set the temperature to 40°C, but the monitor’s temperature keeps rising and stays above 100°C, fluctuating continuously.

The instrument controls the thyristor, and the thyristor controls the heating. Even though your instrument is set to 40°C, it’s not responding to the setting—this likely indicates that the thyristor has failed.

The customer sets the temperature to 90; it may rise back up to between 100 and 110. If set to 100, it could climb above 120 degrees.

The heating rate does not match the PID control parameters; parameter tuning is required.

Press the green button, and the stove heats up instantly—very quickly. Once you press it, the temperature rises rapidly; you must turn it off right away.

The old‑style power regulator suffered a controlled spark‑over breakdown, so it was replaced with a new model featuring a white housing.

Now, during the heating process, once the temperature reaches two or three hundred degrees, it no longer rises, and then the cooling phase begins.

Inspect and replace the trigger wire (the trigger wire between the instrument and the controllable regulator).

There’s an issue with the temperature on our furnace: we’ve set it to 750°C, but half of the furnace heats up normally while the other half only reaches 670°C. What could be the problem?

A broken connection at the furnace outlet terminal in the temperature zone causes no heating and temperature deviation; reconnect the disconnected heating wire.

The tube furnace temperature is rather unstable, and the heating ramp exhibits significant overshoot.

The parameter was modified; it has been restored.

The temperature-programming sequence for the tube furnace is still running, but the temperature has already dropped.

Instrument parameters have been modified: Inp “1” has been changed to “6”, Inp is set to 1, OPH is set to 50°C for heating, and OPH is set to 70°C for heating; however, the target value cannot be reached. Replace the rods (φ14×150×110), 2 pieces.