The display is adjustable, but it does not heat up, and the power‑out indicator light remains off.

It is currently in the operating state, but the temperature is not rising, the output indicator light is off, and there is no temperature inside the furnace tube. A signal cable connector was loose, causing poor contact. Tightening it resolved the issue.

Does not heat up

Communicated with the customer that the fuse had blown and recommended purchasing a new one for replacement.

BTF-1100C-S does not heat up.

The furnace lining material is damaged and needs to be returned to the factory for repair; once the warranty has expired, the costs will be borne by the customer.

The forced-air drying oven is not heating up.

The equipment’s heating element has burned out; replacement is recommended.

Cannot heat or cool anymore.

A wire break was detected at the terminal block; students have been instructed to perform the installation.

Does not heat up

The equipment’s transformer has burned out and needs to be replaced.

The stove is not heating up.

The customer opened the circuit themselves for inspection and found that a wire in the heating section had come loose. After reconnecting it, the heating function operated normally.

The tube furnace in our lab is not heating properly; the power‑input indicator does not flash, and even after clicking “Turn On,” the indicator still fails to flash, preventing normal heating.

Temperature parameter error; correcting the temperature rise restores normal operation.

The tube furnace temperature has remained at 51°C, and it is not rising.

1. The connecting nut between the heating rod and the aluminum finned tube became loose and melted. 2. The nuts between the heating rods were loose. 3. Instrument parameters had been altered (the issue has been resolved, and the equipment is now operating normally).

BTF-1100C-S: This newly arrived tube furnace exhibited a significantly higher PV temperature than the SV temperature during its first no-load firing.

Heating rod broken