The circuit breaker trips during temperature regulation, and the temperature fails to rise. Additionally, at 900°C, the circuit breaker trips again; however, upon inspection, the fuse is intact.

1. Heating rod broken 2. Increase the power output slightly

The device trips upon power-up.

At Kunming University of Science and Technology, the overheating trip has been resolved; the circuit breaker was undersized.

The switching power supply trips automatically.

The thermocouple wiring is loose; inspect and tighten the connection.

The stove tripped the circuit breaker, and even after resetting it, there’s still no power—looks like a short circuit caused the automatic trip.

The customer’s power switch has tripped (requiring maintenance and verification of the circuit breaker’s capacity), which is unrelated to the equipment.

The circuit breaker on the equipment has tripped.

Overheating of the transformer leads to insulation degradation and short circuits. When the current is high, replace the transformer with one of the same rating; alternatively, install additional fans or modify the rear cover. During operation, add fans to enhance heat dissipation.

The device trips when powered on.

Loose wiring at the input and output terminals of the AC contactor caused excessive current and overheating during startup, resulting in a trip; the thyristor also experienced breakdown and short circuit. Replace the thyristor (25 A – 1600 V).

1. The room temperature often fails to rise, 2. When the temperature does increase, it frequently trips the circuit breaker, 3. At high temperatures, the heating sometimes stops altogether.

Power regulator failure (for older‑model power regulators): If necessary, replace the temperature‑control circuit and the current controller. The customer’s mains switch is rated at 16 A; we recommend upgrading to a 32 A or 40 A unit.

Both furnaces are currently experiencing issues: the 1,500°C furnace has an internal electrical leak, and a student was electrocuted. Additionally, the 1,700°C furnace trips the circuit breaker whenever it is operated.

The MF-1500C-S customer connected the power cable backward (it has since been corrected), and the OPL value on the MF-1700C-S device was changed from 48 to 25.

After powering on and initiating heating, the circuit breaker trips. Currently, only the control panel is functioning; the heating button is unresponsive, and the “turn on” command has no effect—no heating occurs.

The issue was caused by the customer’s connection: 220V was mistakenly connected to a 380V supply. Replace one AC contactor (NXC-32 or CJX2-3210) and test again.

The MF-1700C-S furnace has been experiencing frequent circuit breaker trips at the customer’s site.

The heating rod screw was loose; after tightening it, no abnormalities have been observed so far.

Equipment trip

Video inspection: Instrument parameters were modified. After restoration, no abnormalities were detected. The customer has been trained.