The plasma source has a blue screen upon startup, and no numbers are displayed.

The RF power supply is defective and needs to be returned to the factory for repair.

The GMF-2Z touch screen is not lighting up.

When troubleshooting electrical issues with a multimeter, focus on the switching power supply and verify that the 24 V output is within specification.

The sublimation unit’s touch screen is not displaying. Verify the input voltage: 12 V (normal is 24 V).

Check the input voltage: 12V (normal is 24V)

A short circuit occurred during operation, and now the screen is black.

The fuse had blown; after replacing it with a new one, the device is functioning normally. No short circuits were detected during the inspection, so we’ve asked the customer to monitor its operation.

The tube furnace’s display is flickering.

Just reconnect the thermocouple, and that’ll do.