The flow meter is out of control: at a setpoint of 2000 sccm, the display reads over 400, and the reading resets to zero when the gas supply is turned off.

1. Zero drift 2. Adjust the zero point on-site with the customer via video 3. An anomaly has reoccurred; the flowmeter has been sent back for zero calibration.

Air fryer pipe

The flange vent is suspected to be blocked. Clean the surface of any dirt, inspect it for blockages, and address the issue accordingly.

The pressure gauge does not return to zero when there is no air supply.

An operational error: the shut-off valve on the gas outlet of the gas distribution cabinet, leading to the left-hand flange, was not opened. Please open the aforementioned shut-off valve and reset the pressure gauge to zero.

The equipment’s vent valve is open but no air is flowing.

The customer connected the intake pipe incorrectly; replace the outlet.

Right now, the CVD ventilation isn’t working.

Upon inspection, the equipment flanges were found to be fault-free; the customer conducted a self‑inspection and identified an issue with the gas supply pipeline.

The gas line is not supplying gas.

The ±15 V switching power supply is faulty (replace it with a D-30F15 ±15 V unit of the same specifications).

The set value was 300, but the actual tidal volume was zero.

The flow meter is out of control, the gas‑path instruments are malfunctioning, and flow regulation is intermittent. The instrument’s microchip has failed.

The customer reported that the pressure gauge on the gas supply system is displaying an incorrect reading.

In the GZM-3Z gas‑pipeline cabinet, the pressure gauge needle remains at 0.06 MPa under atmospheric pressure and fails to return to zero; replace it with a gauge of the same specification.

BTF-1200C-PECVD-500A: The flow readings for the first and third gas lines are already out of range, and this occurs regardless of whether gas is flowing or not.

The 5V input and output of the measuring instrument are functioning normally. The touchscreen program is refreshing.