undefined
+
  • undefined

1200℃ Continuous Feed/Discharge Rotary CVD Growth System BTF-1200C-II-R-CVD

Equipment Model: BTF-1200C-II-R-CVD

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Key Features

    · The furnace tube can rotate 360 degrees, and quartz baffles on the inner wall facilitate powder turnover, promoting more uniform sintering.

    ·   The entire furnace, along with the feeding and discharging systems, can tilt as a unit, facilitating material discharge; the tilt angle is adjustable between 0° and 35°.

    ·   The feeding system features continuously adjustable feed speed, and when feeding is stopped and the furnace is tilted to its maximum angle, no material leaks out of the system. The feeding system is easily disassembled; after disassembly, manual loading can be performed, allowing the equipment to operate normally.

    ·   The furnace tube rotates steadily, without any vertical or lateral oscillation; the rotational speed is continuously adjustable from 0 to 60 r/min and can sustain continuous operation at 60 r/min for extended periods (>24 hours).

    Structural Description

    · Heating System: The heating chamber is constructed from alumina fiber materials, whose extremely low thermal conductivity and specific heat capacity ensure rapid furnace temperature rise and minimal heat storage, making it particularly well suited for materials‑processing research in academic and research institutions. The heating elements are resistance wires, ideal for medium‑ to low‑temperature heating applications. The high‑temperature vacuum resistance furnace employs a conventional closed‑loop negative feedback control system; the temperature controller is an intelligent programmable unit that regulates temperature via power modulators. Load control utilizes a low‑voltage, high‑current approach, significantly extending the service life of the heating elements. Temperature measurement is performed with K‑type thermocouples. A single‑loop temperature controller can auto‑tune PID parameters without operator intervention, automatically adjusting the output signal to accommodate varying heating and cooling profiles. Additionally, the system incorporates over‑temperature, under‑temperature, and thermocouple‑open fault alarms, substantially reducing the reliance on operator expertise.

    · Feeding System: The feed system features continuously adjustable speed. When feeding is stopped and the furnace is tilted to its maximum angle, no material will leak out of the system. The feed system is easily disassembled, allowing manual loading after removal, ensuring normal equipment operation.

    · Discharge system: Automatically discharges material when the furnace body is tilted for discharge.

    ·   Gas Control System: This gas‑control system can blend two gases in specified flow ratios, or operate them independently. The mass flow meters are mounted within a sealed, mobile cabinet and are connected via ultra‑clean, double‑polished stainless‑steel tubing fitted with precision double‑ferrule fittings.

    ·   Low-vacuum system: This system is equipped with a mechanical vacuum pump, enabling vacuum pumping and allowing the equipment to carry out reactions under vacuum conditions.

    ·   Furnace tube rotation system: Enables the furnace tube to rotate uniformly at 360°, with a continuously adjustable speed ranging from 0 to 60 r/min, and capable of sustained operation at 60 r/min.

    Heating system

    · Rated voltage: AC 220V   50/60 Hz, Rated power: 4 kW

    ·   Maximum temperature: 1200℃

    ·   Operating temperature: ≤1100℃

    · Heating zone length: 200 + 200 mm, constant-temperature zone length: 250 mm

    · Furnace tube specifications: custom-shaped quartz tubes, Φ60×420 mm (at both ends) + Φ100×360 mm (in the middle)

    · Furnace tube rotation speed: 0–60 r/min

    · Furnace tilt angle: 0–35°

    · Temperature control method: 30‑segment programmable temperature control, with auto‑tuning of PID parameters; the operating interface is a 10‑inch LCD touch screen.

    · Temperature control accuracy: ±1℃

    · Heating element: resistance wire

    · Thermocouple type: Type K

      

    Gas Control System

    · Standard range (N 2 ): Route 1 200 sccm, second channel 500 sccm, third channel with selectable range

    · Accuracy: ±1.5% F.S.

    · Linearity: ±1% F.S.

    · Repeatability: ±0.2% F.S.

    · Response time: for gas characteristics, 1–4 seconds; for electrical characteristics, 10 seconds.

    · Working pressure differential range: 0.1–0.5 MPa

    · Maximum pressure: 3 MPa

    · Interface: Φ6, 1/4''

    · Display: 4-digit display

    · Operating ambient temperature: 5–45°C, high-purity gas

    · Pressure vacuum gauge: −0.1 to 0.15 MPa, 0.01 MPa per division

    · Gate valve: Φ6

    · Double-sided polished stainless steel pipe: Φ6

    · Connector Type: Double Ferrule Stainless Steel Connector

    Low-vacuum system

    · Power supply: 220V ±10%, 50–60 Hz; Power: 1 kW

    · Operating ambient temperature: 5~ 40℃

    · Relative air humidity: ≤85%

    · Pumping speed: 10 m³/h

    · Ultimate vacuum level: 5x10 -1 Pa

    · Working pressure range: 1.01325X105 ~1.33×10 -2 Pa

    · Withstand voltage: 0.03 MPa

    · Oil capacity: 1.1 L

    · Inlet/Outlet diameter: KF25

    · Noise: 50 dB

    ·   Connection method: A bellows is used, and the manual damper valve is connected to the bellows.

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Consumable parts such as thermocouples, sealing rings, and furnace tubes are not included.

    · Damage caused by human error or by the use of corrosive and acidic gases is not covered under warranty.

    Precautions

    ·   The internal air pressure must not exceed 0.02 MPa (gauge pressure).

    ·   For experiments involving sample heating, it is not recommended to operate the system with both the vacuum valve and the purge valve at the furnace tube flange closed. If it is necessary to close these valves during heating, continuously monitor the pressure gauge; if the pressure exceeds 0.02 MPa, immediately open the vent valve to prevent accidents such as rupture of the furnace tube or ejection of the flange.

    ·   If, due to process requirements, flammable, explosive, or toxic gases must be used, you are responsible for implementing all necessary safety and explosion‑proof measures. Our company shall not be liable for any issues arising from improper operation during the use of such gases.

    ·   As the furnace is used over time, minor cracks may develop; this is normal, and the cracks can be repaired with an alumina coating.

Consultation Message