undefined
+
  • undefined

1700℃ High-Temperature CVD System BTF-1700C-CVD

Equipment Model: BTF-1700C-CVD Equipment Overview: This system comprises a 1700°C tube furnace, a three‑channel mass flow meter system, and a domestically produced high‑vacuum system (with optional high‑ and low‑vacuum configurations). The equipment features mature technology, reliable performance, uniform temperature distribution, and a well‑designed structure. Its gas‑delivery system is controlled via an LCD touch‑screen interface, ensuring exceptionally simple and convenient operation. All technical parameters are clearly displayed, and the 30‑segment programmable temperature control system provides instant visibility, giving you full oversight of the entire experimental process. Process Applications: Primarily used in high‑temperature experimental environments such as vacuum or atmosphere sintering and substrate coating.

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Main Parameters

    · Voltage: AC 208-240V Single Phase, 50/60 Hz

    ·   Rated power: 5.2 kW

    ·   Maximum temperature: 1650℃

    ·   Operating temperature: ≤1550℃

    · Recommended heating rates: ≤10°C/min below 1400°C; ≤5°C/min from 1400°C to 1600°C.

    ·   Heating zone length: 300 mm, Constant-temperature zone length: 150 mm

    · Heating element: silicon molybdenum rod

    Heating furnace

    ·   Insulation material: An inorganic material composed of alumina fibers, formed by vacuum adsorption.

    · Structure: Closed-type

    · External dimensions: 610  x 440 x 720 mm (Length × Width × Height)

    Furnace tube &Flange

    ·   Furnace tube specifications: Corundum tube Φ60/80 x 1000 mm

    ·   The furnace tube is equipped at both ends with stainless steel sealing flanges, including precision needle valves, pointer-type vacuum pressure gauges, and hose‑type swan‑neck fittings; the ports can be expanded to suit specific applications.

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control allows customers to customize heating and cooling profiles according to their specific needs.

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Standard accessories

    · One high-purity corundum tube; 1 set of stainless steel sealing flanges; 2 sets of fluororubber sealing rings; 4 alumina tube plugs; 1 pair of high-temperature gloves; 1 crucible hook; 1 small fuse.

    Gas‑line system parameters (mass flow meters available in multiple channel configurations)

    · Reference flow range of the mass flow meter (N 2 ): 100 sccm, 200 sccm, 500 sccm, 1000 sccm (range selectable)

    · Accuracy: ±1.5%

    · Linearity: ±0.5–1.5%

    · Repeatability: ±0.2%

    · Response time: ≤10 sec

    · Working pressure differential range: 0.1–0.5 MPa

    · Maximum pressure: 3 MPa

    · Interface: Φ6, 1/4''

    · Display: 4-digit display

    · Operating ambient temperature: 5–45°C, high-purity gas

    · Pressure vacuum gauge: −0.1 to 0.15 MPa, 0.01 MPa/division

    · Gate valve: Φ6

    · Double-sided polished stainless steel pipe: Φ6

    · Connector Type: Double Ferrule Stainless Steel Connector

    Low-vacuum system parameters

    · Power supply: 220V, 50Hz; Motor power: 0.4 kW

    · Pumping speed: 10 m³/h

    · Ultimate vacuum level: 5x10 -1 Pa

    · Inlet/Outlet Connection: KF25

    · Oil capacity: 1.1 L

    · Motor speed: 1440 rpm

    · Operating ambient temperature: 5~ 40℃

    · Noise: ≤56dB

    Resistance vacuum gauge parameters

    · Regulatory requirements: Resistance gauge

    · Measurement range: 10 5 Approximately 10 -1 Pa

    · Measurement channel: 1 channel

    · Control range: 5×10 3 ~5×10 -1 Pa

    · Control method: Relay contact output, load capacity AC 220 V   3A (or DC 28V)   3A) Resistive Load

    · Power supply: 90–260 V   50Hz or 220V±10% 50Hz

    · Power consumption: 20W

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Consumable parts such as thermocouples, sealing rings, and furnace tubes are not included.

    · Damage caused by human error or by the use of corrosive and acidic gases is not covered under warranty.

    Precautions

    ·   The internal air pressure must not exceed 0.02 MPa (gauge pressure).

    ·   Due to the high internal pressure of the gas cylinder, a pressure-reducing valve must be installed when introducing gas into the quartz tube. To ensure safety, it is recommended to use a pressure-reducing valve with a rated pressure below 0.02 MPa.

    ·   For experiments involving sample heating, it is not recommended to operate the system with both the vacuum valve and the inlet valve at the furnace tube flange closed. If it becomes necessary to close these valves during heating, continuously monitor the pressure gauge; if the pressure exceeds 0.02 MPa, immediately open the vent valve to prevent potential accidents, such as rupture of the furnace tube or ejection of the flange.

    ·   Customers are advised against using flammable, explosive, or toxic gases. If such gases are unavoidably required for process reasons, please implement all necessary safety and explosion‑proof measures on your own. Our company shall not be held liable for any issues arising from the use of flammable, explosive, or toxic gases.

Related products