undefined
undefined
+
  • undefined
  • undefined

1200℃ High-Temperature, High-Pressure Vacuum Tube Furnace BTF-1200C-HP

Equipment Model: BTF-1200C-HP Equipment Overview: The BTF-1200C-HP is a high‑temperature, high‑pressure tube furnace. Its furnace tube is made of nickel‑based alloy steel, capable of withstanding temperatures up to 1100°C. Under operating conditions at 800°C, the maximum pressure can reach 4 MPa. The sealing system employs CF flange connections, with metal oxygen‑free copper gaskets. This model also features excellent oxidation resistance in its furnace tube. The equipment is equipped with a pressure sensor and a solenoid valve; when the pressure sensor detects an internal pressure exceeding the set value, it sends a signal to the high‑pressure solenoid valve, which opens to release excess pressure, thereby maintaining the internal pressure at the preset level.

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Main Parameters

    · Voltage: AC 208-240V, Single Phase, 50/60Hz , Rated power: 3 kW

    · Maximum temperature: 1200℃

    · Continuous Operating temperature: ≤1100℃

    · Recommended heating rate: ≤15°C/min

    ·   Heating zone length: 440 mm, Constant-temperature zone length: 220 mm

    · Heating element: resistance wire

    · Thermocouple: Type K thermocouple

    Furnace tube &Flange

    · Furnace tube material: nickel-based alloy steel

    · Furnace tube diameter: Φ60×1000 mm (customization available)

    ·   CF flange sealing; the sealing ring is made of oxygen-free copper.

    ·   Equipped with a pressure sensor and a solenoid valve.

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control allows customers to customize heating and cooling profiles according to their specific requirements.

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Equipment dimensions

    · Device dimensions: 590x380x500mm

    Inlet pressure

    · When the instrument is in operation, both pressure and temperature must be taken into account.

    Under positive-pressure conditions

    Pressure

    Temperature

    ≤4Mpa

    ≤800℃

    ≤2.5 MPa

    ≤900℃

    ≤1.2 MPa

    ≤1000℃

    ≤0.8 MPa

    ≤1100℃

    Under atmospheric pressure conditions

    ≤1000℃

    Under negative pressure conditions

    10 -1 Pa (using a mechanical pump) (optional)

    10 -4 Pa (with molecular pump) (optional)

    Enter /Exhaust structure

    · Air intake structure: Each furnace is equipped with an independent… Phi 6 mm ferrule inlet, mechanical pressure gauge, precision shut-off valve

    · Exhaust structure: Each furnace is equipped with an independent… Phi 9.2 mm exhaust port and ball valve

    Expansion port

    ·   This equipment is equipped with a KF25 pumping port and a KF16 vacuum gauge connection, enabling integration with a vacuum system; the base vacuum can reach 0.5 Pa.

    Pressure relief device

    · Equipped with a high-precision electromagnetic pressure relief valve, the system automatically opens the valve to release excess pressure when the chamber pressure exceeds the preset safety threshold, ensuring experimental safety.

    Cooling device

    · Heat sinks are mounted on both ends of the flange, effectively enhancing the device’s heat dissipation efficiency.

    Gas Supply System (Optional)

    · Select one or more gases and configure their mixing ratio according to process requirements.

    · A rotameter or a mass flow meter may be used to control the inlet air flow rate.

    Vacuum System (Optional)

    · A vacuum assembly or a vacuum unit can be selected to match the equipment for vacuum pumping.

    · Pumping rate: 10 m³/h

    · Vacuum gauge measurement range: 10 5 Approximately 10 -1 Pa

    · Ultimate pressure: 5×10 -1 Pa (cold, no-load)

    Security Protection

    · This device is equipped with over-temperature and thermocouple failure alarms.

    · The device can withstand positive pressure within its rated range; if the set pressure threshold is exceeded, it automatically vents to ensure safety.

    Warranty period

    · Host One-year warranty, lifetime maintenance.

    · Upon delivery of the equipment, please contact our company to arrange for a dedicated after-sales technician to visit your site for installation, commissioning, and training.

    ·   Consumable parts such as metal tubes, thermocouples, and sealing rings are not covered under the warranty.

    ·   Damage caused by process-related issues, such as the use of corrosive gases or strong acids and bases, is not covered under warranty.

    Precautions

    ·   This equipment uses metal tubing; samples with acidic or alkaline properties are not recommended.

Consultation Message