1200℃ Three-Zone Sliding-Rail Rapid Heating and Cooling Furnace BTF-1200C-III-SL

Equipment Model: BTF-1200C-III-SL Equipment Overview: The BTF-1200C-III-SL temperature-zone slide‑rail furnace is a modified version of an open‑type vacuum tube furnace. During operation, the desired constant‑temperature zone can be precisely positioned directly at the sample, enabling rapid heating. Conversely, the high‑temperature zone can be swiftly moved away from the sample, exposing it to ambient conditions for rapid cooling. This specialized furnace is designed specifically for the growth of two‑dimensional materials and is also well suited for experiments requiring fast heating and cooling rates.

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Key Features

    ·   It can be evacuated, and it can also be purged with an inert gas.

    ·   The constant-temperature field required for the experiment can be directly delivered to the sample, enabling rapid heating.

    ·   The temperature field in the experimental zone can be directly displaced away from the sample, enabling rapid cooling of the sample.

    ·   The clamp flange eliminates the cumbersome procedures involved in loading and unloading materials.

    ·   The furnace chamber is lined with high-purity alumina ceramic fiber, which minimizes energy loss to the greatest extent.

    ·   The pneumatic spring‑assisted structure enables the operator to open it safely at a wide angle.

    Main Parameters

    · Voltage: AC 208–240 V; frequency 50/60 Hz (standard power cord length: 2.5 m)

    ·   Rated power: 7 kW

    · Maximum temperature: 1200℃

    ·   Operating temperature: ≤1100℃

    · Recommended heating rate: ≤20°C/min

    ·   Heating zone length: 300+300+300 mm, constant-temperature zone length: 150+150+150 mm

    · Heating element: resistance wire

    · Thermocouple: Type K thermocouple

    Furnace tube &Flange

    ·   Furnace tube specifications: High-purity quartz tube, Φ50/60/80/100/130 × 2300 mm (tube diameter customizable)

    ·   Stainless steel sealing flange ( Includes precision needle valves, dial-type mechanical pressure gauges, and hose‑type swivel fittings, sealed with fluororubber O‑rings.

    · It is equipped with a mechanical pressure gauge having a measurement range of -0.1 to 0.15 MPa.

    · The flange is equipped with a corresponding interface, allowing customers to connect the appropriate equipment as needed.

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control allows customers to customize heating and cooling profiles to meet diverse process requirements.

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Size &Weight

    · Stove dimensions: 1080  x 440   x 590mm

    · Stove + sliding rails: 2000   x 490   x 760mm

    · Equipment net weight: approximately 195 kg

    Vacuum System (Optional)

    ·   Model: Baosi Vacuum DRV-10

    · Pumping speed: 10 m³/s 3/h

    · Ultimate pressure: 5×10 -1 Pa (no-load, cold state)

    · Motor power: 550 W

    · Weight: 25 kg

    · Required oil quantity: 1.1 L

    (Optional: domestically produced or imported high-vacuum unit)

      

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Consumable items such as thermocouples, fluororubber rings, quartz tubes, and quartz boats are not included.

    · Issues arising from improper use or the use of corrosive or acidic gases are not covered under the warranty.

    Precautions

    ·   The gas pressure inside the quartz tube must not exceed 0.02 MPa (gauge pressure).

    ·   For experiments involving sample heating, it is not recommended to close the vacuum and inlet valves at the flange end of the furnace tube. If it becomes necessary to close these valves during sample heating, continuously monitor the mechanical pressure gauge; if the gauge reading exceeds 0.02 MPa, immediately open the vent valve to prevent accidents such as furnace tube rupture or flange ejection.

    ·   If, due to process requirements, flammable, explosive, or toxic gases must be used, you are responsible for implementing all necessary safety and explosion‑proof measures. Our company shall not be liable for any issues arising from improper operation during the use of such gases.

    · As the furnace is used over time, minor cracks may develop in the furnace chamber; this is normal, and the cracks can be repaired with a high-purity alumina repair coating.