undefined
undefined
+
  • undefined
  • undefined

1400°C Three-Zone Vacuum High-Temperature Tube Furnace BTF-1400C-III

Equipment Model: BTF-1400C-III Equipment Overview: The BTF-1400C-III is a CE‑certified high‑temperature tube furnace with a maximum operating temperature of 1400°C and a rated power of 13 kW. It features three independently controllable heating zones, allowing users to tailor the temperature profile to experimental needs—either by configuring three distinct temperature zones or by creating an exceptionally long constant‑temperature zone. Equipped with stainless steel flanges, it offers quick and convenient installation and removal. This equipment boasts mature technology, reliable quality, uniform temperature distribution, and a well‑designed structure. Application Areas: Widely used in CVD experiments, phosphor synthesis, vacuum or atmosphere sintering, substrate coating, and other applications requiring elevated heating temperatures. It is one of the ideal pieces of equipment indispensable for leading materials research laboratories.

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Key Features

    ·   Silicon-carbide rod heaters are densely arranged throughout the furnace chamber, ensuring a uniform temperature field and an exceptionally long constant-temperature zone.

    ·   Double-layer shell structure with an air-cooling system

    ·   Well-structured, with simple and convenient experimental procedures.

    ·   Install flange support brackets to reduce the load on the furnace body.

    ·   High-purity alumina is used as the furnace lining material, and the furnace surface is coated with a high-temperature alumina coating imported from the United States, which enhances heating efficiency and extends the service life of the furnace.

    Main Parameters

    · Voltage: 380 V, Rated Power: 13 kW

    · Maximum temperature: 1400℃

    · Continuous Operating temperature: ≤1300℃

    · Recommended heating rates: ≤10°C/min below 1200°C; ≤5°C/min from 1200°C to 1300°C.

    ·   Heating zone length: 180 + 220 + 180 mm; constant-temperature zone length: 450 mm.

    · Heating element: silicon carbide rod

    · Thermocouple: Type S thermocouple

    Furnace tube &Flange

    ·   Furnace tube diameter: Alumina tube Φ60/Φ80 × 1200 mm (tube diameter customizable)

    ·   It is equipped with a set of stainless steel quick‑connect flanges (available in two types: clamp flanges and hinged flanges) and sealed with silicone O‑rings.

    · Vacuum level: 0.5 Pa (using a mechanical pump)

    · A mechanical pressure gauge is installed, with a measurement range of -0.1 to 0.15 MPa.

    · Use a pipe plug to block the thermal radiation emitted from the center of the pipe; this must be done before heating and positioned correctly within the furnace chamber.

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control allows customers to customize heating and cooling profiles according to their specific requirements.

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Vacuum System (Optional)

    ·   Model: DRV-10

    · Pumping rate: 10 m³/h

    · Ultimate pressure: ≤5×10 -1 Pa (no-load, cold state)

    · Oil capacity: 1.1 L

    · Inlet/Outlet Port Diameter: KF25

    (Optional: domestically produced or imported high-vacuum unit)

      

    Size &Weight

    · Device dimensions: 61 0 x 440 x 720mm

    · Equipment net weight: approximately 120 kg

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Consumable items such as thermocouples, silicone rings, alumina tubes, and sample crucibles are not included.

    · Damage caused by improper use or by exposure to corrosive or acidic gases is not covered under the warranty.

    Precautions

    ·   The internal air pressure must not exceed 0.02 MPa (gauge pressure).

    ·   For experiments involving sample heating, it is not recommended to operate the system with both the exhaust valve and the inlet valve at the furnace tube flange closed. If it becomes necessary to close these valves during heating, closely monitor the pressure gauge; if the pressure exceeds 0.02 MPa, immediately open the vent valve to prevent potential accidents, such as rupture of the furnace tube or ejection of the flange.

    ·   If, due to process requirements, flammable, explosive, or toxic gases must be used, you are responsible for implementing all necessary safety and explosion‑proof measures. Our company shall not be liable for any issues arising from improper operation during the use of such gases.

    · Due to the manufacturing process of alumina tubes, there is a risk of tube breakage during the sintering process, which cannot be entirely eliminated. Please be advised.