Double-layer sliding-rail furnace BTF-1200C-II-SL-2D-CVD

Equipment Model: BTF-1200C-II-SL-2D-CVD

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Structural Features

    · The furnace tube is a quartz tube with an outer diameter of Phi 25*1400mm

    · The left and right flanges are lateral clamp-type flanges.

    · The left-hand flange is equipped with one inlet port controlled by a shut-off valve and one pressure gauge.

    · The right-hand flange is equipped with a PCG500 vacuum gauge, a KF25 vacuum inlet, and a ball-valve‑controlled exhaust port.

    · The slide rail is equipped with a dual-zone heating furnace, with the furnace body arranged in two tiers—one above the other.

    · The equipment features a load-bearing frame‑based chassis, with the heating furnace mounted on a sliding rail mechanism that enables rapid left–right heating and cooling via guided rails.

    · A pneumatic cabinet and two vacuum pumps are installed beneath the guide rail.

    · The device features a stylish high-gloss white finish, combining minimalist aesthetics with a sense of sophistication, while delivering intuitive, user-friendly operation and precise temperature control.

    Main Parameters

    · Voltage: 220V, Power: 7 KW

    ·   Maximum temperature: 1200℃

    · Operating temperature: ≤1100℃

    · Recommended heating rate: ≤10°C/min

    · Furnace tube specifications: high-purity quartz tube, Φ25 × 1400 mm

    · Heating zone length: 200 + 200 mm (2 units)

    · Heating element: resistance wire

    · Thermocouple type: Type K

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Gas pipeline cabinet

    · Gas flow rate for line 3 (first layer receives a single stream; second layer uses a 2-in-1 configuration)

    Vacuum assembly

    · Vacuum pump DRV16 + vacuum gauge PCG550

    Equipment dimensions &Weight

    · Overall equipment dimensions: 2090 x 700 x 1680 mm (L x W x H)

    · Total equipment weight: approximately 310 kg

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Consumable items such as thermocouples, sealing rings, furnace tubes, and sample holders are not included.

    · Damage caused by improper user handling or process-related issues—such as exposure to corrosive gases, strong acids, or strong bases—is not covered under the warranty.

    Precautions

    ·   The gas pressure inside the quartz tube must not exceed 0.02 MPa (gauge pressure).

    ·   For experiments involving sample heating, it is not recommended to operate the system with both the vacuum valve and the purge valve at the furnace tube flange closed. If it is necessary to close these valves during heating, continuously monitor the pressure gauge; if the pressure exceeds 0.02 MPa, immediately open the vent valve to prevent accidents such as rupture of the furnace tube or ejection of the flange.

    ·   If, due to process requirements, flammable, explosive, or toxic gases must be used, you are responsible for implementing all necessary safety and explosion‑proof measures. Our company shall not be liable for any issues arising from improper operation during the use of such gases.

    · As the furnace is used over time, minor cracks may develop; this is normal, and the cracks can be repaired with an alumina coating.