undefined
+
  • undefined

Vacuum Sublimation Instrument BOF-7-120

Equipment Model: BOF-7-120

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Structural Overview

    · Electrical Control and Furnace Chamber

    The electrical control system employs a combination of temperature‑control instruments and a touchscreen, offering precise temperature regulation and intuitive, user‑friendly operation. Both the instruments and the touchscreen are sourced from renowned domestic brands, ensuring high accuracy and reliable performance.

    The furnace features seven temperature zones with lengths of 100 mm, 300 mm, 100 mm, 150 mm, 150 mm, 100 mm, and 100 mm, respectively. The partition thickness between zones is 10 mm. Each of the seven zones is equipped with 14-point temperature control, can be opened electrically, and includes cooling fans for faster temperature reduction. It is fitted with a 120 mm‑diameter furnace tube, providing greater loading capacity. Heating is achieved via resistance wire, which offers exceptional oxidation resistance. The insulation material is an environmentally friendly, energy‑saving inorganic alumina fiber product formed by vacuum adsorption.

    · Vacuum piping and flanges

    Left-end flange: Features a sliding design for easy and convenient removal and cleaning of the furnace tube. A PTFE gasket is used to seal the gap between the furnace tube and the flange, minimizing contamination sources. Additionally, it is equipped with an observation window and a pre‑installed KF16 vacuum port.

    Right‑hand flange: Features a simplified design; the gap between the furnace tube and the flange is sealed with a PTFE gasket to minimize contamination sources. It is equipped with a bypass bleed valve and a vacuum gauge interface.

    Vacuum piping: The CF100 port on the right‑hand flange is connected to a pneumatic slide valve; downstream of the valve is a cooling and filtration chamber equipped with both a molecular pump and a mechanical pump. A static filter screen is installed between the right‑hand flange and the slide valve assembly. All pipeline connections employ a metal‑to‑metal seal combined with an FKM O‑ring seal to enhance vacuum performance.

    · Traffic control

    This equipment is equipped with a single‑channel mass flow meter, model Qixing D07‑26C, calibrated for nitrogen. The flow range is 5 slm, and the instrument is from a well‑known domestic brand, ensuring high accuracy and reliability. After passing through the flow meter, the gas enters the furnace tube from both the left and right ends, resulting in a relatively stable gas flow that minimizes its impact on the sublimation material.

    · Appearance

    This equipment features a metal‑cased, load‑bearing structure and is finished in a custom orange‑patterned white. Its design is clean and streamlined, offering intuitive, user‑friendly operation, precise temperature control, and superior vacuum performance.

    Main Parameters

    · Rated voltage: AC 380V, Rated power: 12 kW

    · Set the maximum temperature of the equipment to 600℃.

    · Maximum heating rate: 10°C/min

    ·   Furnace tube specifications: outer diameter Φ120 × 1400 × 4 mm

    · Heating zone length: 100+300+100+150+150+100+100 mm

    · Heating element: resistance wire

    · Vacuum break: pressure 0.1–0.3 MPa

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Equipment dimensions &Weight

    · Device dimensions: 1930  x 1900   x   1600 mm (length × width × height)

    · Equipment net weight: approximately 200 kg

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Consumable parts such as heating elements, sealing rings, furnace tubes, and sample holders are not included.

    · Issues arising from improper user handling or the use of corrosive gases, acidic gases, or similar substances are not covered under warranty.

    Precautions

    · As the furnace is used over time, minor cracks may develop in the furnace chamber; this is normal, and the cracks can be repaired with a high-purity alumina repair coating.

Consultation Message