High-Throughput Parallel CVD BTF-1200C-VI-SL-CVD-D280

Equipment Model: BTF-1200C-VI-SL-CVD-D280

Keywords:

Close
  • 产品对比
Total X Clear all
  • Product Information
  • Product Video
  • Experimental Case
  • Accessories and Consumables
  • Structural Description

    · Open‑type heating furnace: The furnace chamber is lined with ultra‑light alumina fiber insulation and equipped with K‑type thermocouples, enabling independent eight‑point temperature control in both the upper and lower sections. The unit offers flexible heating and cooling control, and its open design facilitates convenient replacement and maintenance of the reaction tubes.

    · Open‑type preheating furnace: The heating chamber is constructed from ultra‑light alumina fiber, equipped with K‑type thermocouples and features independent four‑point temperature control for both the upper and lower sections. The unit offers flexible heating and cooling control, and its open design allows for convenient replacement and maintenance of the reaction tubes.

    ·   Gas‑supply system: Enables precise control of multiple gas streams, allowing them to be mixed and proportioned at varying flow rates, while also supporting integrated PLC‑based interlock control.

    · Reaction tube assembly: Uses a quartz tube with water-cooled flanges at both ends and quick-release clamps for easy disassembly, maintenance, and cleaning.

    ·   Intelligent Control System: Employs intelligent PID fuzzy control with 30 programmable segments. Features over-temperature and thermocouple‑open‑circuit alarm functions.

    Main Parameters

    · Voltage: AC 380V, Total heating power: ≤27 KW

    ·   Maximum temperature: 1200℃

    · Use Temperature: 100–1100°C

    · Recommended heating rate: 1–10°C/min

    ·   Heating zone length: 100+200+100+200+200+200 mm (six independent temperature zones)

    · Furnace tube specifications: High-purity quartz tube (outer diameter) 280 mm

    · Heating element: resistance wire

    · Temperature sensing element: Type K thermocouple

    Temperature control system

    ·   30 programmable temperature control segments, with auto-tuning of PID parameters.

    · The operator interface features a 10-inch touchscreen with a built-in PLC control program.

    ·   Modular automated control integrates functions such as heating, intake‑air flow rate setting, vacuum pump on/off, vacuum gauge reading display, chamber pressure setpoint and feedback, and furnace door opening/closing. Parameters can be configured, logged, and operated with a single touch on the touchscreen.

    · Temperature control accuracy: ±1℃

    Flowmeter Control System

    · Gas supply types: 6 varieties

    ·   Standard flow ranges: 20 SLM for argon, 2 SLM for an argon–oxygen mixture (O₂ at 0.1%), and 2 SLM for oxygen; the remaining three channels are calibrated to the argon range, and the measurement range can be adjusted according to the process requirements.

    ·   Accuracy: ±1.0% F.S.

    · Linearity: ±0.5% F.S.;

    · Repeatability: ±0.2% F.S.

    ·   Operating temperature: 4–45°C

    · Working pressure differential: 0.1–0.5 MPa

    · Maximum pressure: 3 MPa

    Vacuum system

    Vacuum pump

    · Power supply: 380 V, 50 Hz; Power: 2.2 kW

    · Pumping speed: 60 m³/h

    · Experimental vacuum level: 1 Pa

    · Inlet diameter: KF40

    · Exhaust port diameter: KF40

    Vacuum gauge

    · Input voltage: 24V 1A DC (with a 220V transformer)

    · Vacuum gauge interface: KF16

    · Digital display: The digital display shows readings in torr.

    · Gas detection: Capable of detecting a wide range of gases, including gas mixtures, with high measurement accuracy and excellent repeatability.

    · Response time: ≤30 ms

    · Measurement range: 3.8×10⁻⁵ — 1125 Torr

    Constant Pressure Control System

    · Structural composition: The system consists of a butterfly valve, a pressure controller, and a vacuum gauge. According to experimental requirements, a butterfly valve (control valve) and a vacuum gauge are installed between the outlet and the pump. The vacuum gauge measures the pressure inside the pipeline, while the PLC adjusts the opening degree of the butterfly valve (control valve) to maintain a constant internal pressure.

    Sliding rail system

    · The furnace body can slide left and right.

    · Sliding method: Electric

    · Sliding speed: 5–10 mm/s

    Cold Trap Recovery System

    · Recycling Structure Design: This unit is equipped with two vacuum cold traps, each connected to the CVD reactor body. It can be internally supplied with a refrigerant—such as alcohol, liquid nitrogen, or cooling water, selected according to the condensation point of the process material—to capture process exhaust gases and excess reaction precursors.

    · Cold trap material: Made of 304 stainless steel.

    · Capacity: 2L

    Chiller

    · Model: CW-5000TG

    · Operating voltage: AC 208–240 V, 50 Hz

    · Operating current: 1.4–2.1 A

    · Compressor power: 0.295 kW

    · Cooling capacity: 2361 BTU/h

    · Water tank capacity: 6L

    · Maximum flow rate: 10 L/min

    · Safety Protection: Compressor overcurrent protection, flow alarm, and overtemperature alarm.

    · Net weight: 26 kg

    · Mechanical dimensions: 55 x 28 x 43 cm

    Size &Weight

    · Device dimensions: 3900  x 1150 x 2000 mm (Length × Width × Height)

    · Equipment weight: approximately 1300 kg

    Precautions

    ·   The internal air pressure must not exceed 0.02 MPa (gauge pressure).

    ·   Due to the high internal pressure of the gas cylinder, a pressure-reducing valve must be installed when introducing gas into the quartz tube. To ensure safety, the pressure after reduction should be maintained at 0.2–0.4 MPa, and the reduced‑pressure gas must pass through a flow meter to regulate the flow before entering the furnace tube.

    ·   For experiments involving sample heating, it is not recommended to close the vacuum and inlet valves at the flange end of the furnace tube. If it becomes necessary to close these valves during sample heating, continuously monitor the mechanical pressure gauge; if the gauge reading exceeds 0.02 MPa, immediately open the vent valve to prevent accidents such as furnace tube rupture or flange ejection.

    ·   If, due to process requirements, flammable, explosive, or toxic gases must be used, you are responsible for implementing all necessary safety and explosion‑proof measures. Our company shall not be liable for any issues arising from improper operation during the use of such gases.

    · As the furnace is used over time, minor cracks may develop in the furnace chamber; this is normal, and the cracks can be repaired with a high-purity alumina repair coating.