    · The equipment is a vertical tube furnace, equipped with upper and lower flanges, with a single-zone heating chamber in the middle.

    · The lower material-feeding platform can move vertically via a motor to pick up the crucible.

    · The furnace body can also be moved vertically, with its slow lifting and lowering achieved by a rear-mounted servo motor paired with a reducer.

    · Both the upper and lower sections of the device are equipped with sliding mechanisms that, in conjunction with built-in motors and lead-screw assemblies, enable vertical movement.

    · A sample stage is fixed to the lower flange for loading and unloading materials; meanwhile, a pressure sensor and a pressure relief port are mounted on the side of the lower flange.

    · The upper flange is equipped with a vacuum gauge and a vacuum port, which are connected to the pump unit via a vacuum pipeline, a manual butterfly valve, and a bellows to achieve vacuum pumping.

    · The furnace tube is an alumina tube with dimensions φ80 × 1200; the middle section of the furnace tube serves as a single-zone heating zone, with a length of 300 mm.

    · The upper flange is equipped with an exhaust port, which is connected via a metal bellows to the side panel of the frame for convenient operator access; the lower flange is fitted with an air inlet.

    · The equipment features a vertical configuration; users can operate it via the side-mounted control panel, which is equipped with a touchscreen and valves, among other components.

    Main Parameters

    · Rated voltage: AC 380V, Power: 8 KW

    ·   Operating temperature: ≤1700℃

    ·   Heating zone length: 300 mm

    · Furnace moving speed: 0.03–3 mm/h

    · Heating element: silicon molybdenum rod

    · Temperature-sensing element: Type B thermocouple

    Furnace tube specifications

    · Corundum tube Φ80*1200mm

    Temperature control system

    ·   Fuzzy PID Control and Self-Tuning Regulation

    ·   Intelligent 30-segment programmable control

    ·   Features ultra‑gentle operation and open‑circuit alarm functionality.

    · Temperature control accuracy: ±1℃

    Equipment dimensions &Weight

    · Vertical furnace: 1200 x 840 x 2750 mm (L x W x H)

    · Pump unit: 710 x 500 x 1500 mm (L × W × H)

    · Total equipment weight: 700 kg for the vertical furnace; 100 kg for the pump unit.

    Warranty period

    ·   One-year warranty, lifetime maintenance.

    · Issues arising from improper user handling or the use of corrosive gases, acidic gases, or similar substances are not covered under the warranty.

    Precautions

    · For experiments involving sample heating, it is not recommended to close the vacuum and inlet valves at the flange end of the furnace tube. If it becomes necessary to close these valves during sample heating, continuously monitor the mechanical pressure gauge; if the gauge reading exceeds 0.02 MPa, immediately open the vent valve to prevent accidents such as furnace tube rupture or flange ejection.

    · If, due to process requirements, flammable, explosive, or toxic gases must be used, you are responsible for implementing all necessary safety and explosion‑proof measures. Our company shall not be liable for any issues arising from improper operation during the use of such gases.

    · Due to the manufacturing process of alumina tubes, there is a risk of tube breakage during the sintering process, which cannot be entirely eliminated. Please be advised.

Consultation Message